A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak

S. Kodaira, K. Morishige, H. Kawashima, H. Kitamura, M. Kurano, Nobuyuki Hasebe, Y. Koguchi, W. Shinozaki, K. Ogura

    Research output: Contribution to journalArticle

    8 Citations (Scopus)

    Abstract

    We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59% of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1% to 1.5% and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.

    Original languageEnglish
    Pages (from-to)129-135
    Number of pages7
    JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
    Volume383
    DOIs
    Publication statusPublished - 2016 Sep 15

    Fingerprint

    linear energy transfer (LET)
    performance tests
    Energy transfer
    Surface properties
    plastics
    Plastics
    Detectors
    sensitivity
    detectors
    Heavy ions
    heavy ions
    ion accelerators
    Krypton
    antioxidants
    krypton
    Antioxidants
    radiation distribution
    Dosimetry
    Particle accelerators
    dosimeters

    Keywords

    • CR-39
    • Dosimetry
    • Heavy ions
    • Nuclear fragmentation
    • Nuclear track detectors
    • Space radiation

    ASJC Scopus subject areas

    • Instrumentation
    • Nuclear and High Energy Physics

    Cite this

    A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak. / Kodaira, S.; Morishige, K.; Kawashima, H.; Kitamura, H.; Kurano, M.; Hasebe, Nobuyuki; Koguchi, Y.; Shinozaki, W.; Ogura, K.

    In: Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms, Vol. 383, 15.09.2016, p. 129-135.

    Research output: Contribution to journalArticle

    Kodaira, S. ; Morishige, K. ; Kawashima, H. ; Kitamura, H. ; Kurano, M. ; Hasebe, Nobuyuki ; Koguchi, Y. ; Shinozaki, W. ; Ogura, K. / A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak. In: Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms. 2016 ; Vol. 383. pp. 129-135.
    @article{699d38a787ed4fbe97df708f21229dab,
    title = "A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak",
    abstract = "We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59{\%} of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1{\%} to 1.5{\%} and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.",
    keywords = "CR-39, Dosimetry, Heavy ions, Nuclear fragmentation, Nuclear track detectors, Space radiation",
    author = "S. Kodaira and K. Morishige and H. Kawashima and H. Kitamura and M. Kurano and Nobuyuki Hasebe and Y. Koguchi and W. Shinozaki and K. Ogura",
    year = "2016",
    month = "9",
    day = "15",
    doi = "10.1016/j.nimb.2016.07.002",
    language = "English",
    volume = "383",
    pages = "129--135",
    journal = "Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms",
    issn = "0168-583X",
    publisher = "Elsevier",

    }

    TY - JOUR

    T1 - A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak

    AU - Kodaira, S.

    AU - Morishige, K.

    AU - Kawashima, H.

    AU - Kitamura, H.

    AU - Kurano, M.

    AU - Hasebe, Nobuyuki

    AU - Koguchi, Y.

    AU - Shinozaki, W.

    AU - Ogura, K.

    PY - 2016/9/15

    Y1 - 2016/9/15

    N2 - We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59% of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1% to 1.5% and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.

    AB - We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59% of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1% to 1.5% and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.

    KW - CR-39

    KW - Dosimetry

    KW - Heavy ions

    KW - Nuclear fragmentation

    KW - Nuclear track detectors

    KW - Space radiation

    UR - http://www.scopus.com/inward/record.url?scp=84978402728&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=84978402728&partnerID=8YFLogxK

    U2 - 10.1016/j.nimb.2016.07.002

    DO - 10.1016/j.nimb.2016.07.002

    M3 - Article

    AN - SCOPUS:84978402728

    VL - 383

    SP - 129

    EP - 135

    JO - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

    JF - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

    SN - 0168-583X

    ER -