A study on a newly designed aortofemoral prosthetic y graft

T. Fujimoto, S. Nakano, H. Iwamura, Y. Shiraishi, T. Yambe, M. Umezu

Research output: Contribution to journalArticle

Abstract

Aortofemoral prosthetic Y grafts are applied for treating abdominal aortic aneurysms. In this study, experiments were conducted in order to demonstrate the effects of an incremental increase in branch diameter in a newly designed aortofemoral Y graft under conditions of both steady and pulsatile flow conditions. It was revealed that in Y grafts, the incremental increase by only 4 mm in the branch diameter affects hydrodynamic characteristics drastically under conditions of both steady and pulsatile flow conditions.

Original languageEnglish
Pages (from-to)59-63
Number of pages5
JournalBiocybernetics and Biomedical Engineering
Volume28
Issue number1
Publication statusPublished - 2008 Dec 1

Keywords

  • Aortofemoral prosthetic y graft
  • Bifurcation
  • Mock circuit
  • Pulsatile flow
  • Steady flow

ASJC Scopus subject areas

  • Biomedical Engineering

Fingerprint Dive into the research topics of 'A study on a newly designed aortofemoral prosthetic y graft'. Together they form a unique fingerprint.