A surgical robot with vision field control for single port endoscopic surgery

Yo Kobayashi, Yu Tomono, Yuta Sekiguchi, Hiroki Watanabe, Kazutaka Toyoda, Kozo Konishi, Morimasa Tomikawa, Satoshi Ieiri, Kazuo Tanoue, Makoto Hashizume, Masaktsu G. Fujie

Research output: Contribution to journalArticle

31 Citations (Scopus)

Abstract

Background Robotic end-effectors for single port endoscopic surgery (SPS) require a manual change of vision field that slows surgery and increases the degrees of freedom (DOFs) of the manipulator. Methods A new surgical prototype robot has dynamic vision field control and a master controller to manipulate the endoscopic view. It uses positioning (4 DOF) and sheath (2 DOF) manipulators for vision field control, and dual tool tissue manipulators (gripping, 5 DOF; cautery, 3 DOF). Results The robot is feasible in vitro. 'Cut and vision field control' (using tool manipulators) was suitable for precise cutting tasks in risky areas; 'cut by vision field control' (using the vision field control manipulator) was effective for rapid macro cutting of tissues. A resection was performed using a combination of both methods. Conclusions The novel robotic system is feasible, but further studies are needed to address its performance in vivo.

Original languageEnglish
Pages (from-to)454-464
Number of pages11
JournalInternational Journal of Medical Robotics and Computer Assisted Surgery
Volume6
Issue number4
DOIs
Publication statusPublished - 2010 Dec

Fingerprint

Surgery
Manipulators
Robotics
Tissue
Cautery
End effectors
Macros
Robotic surgery
Robots
Controllers

Keywords

  • endoscopic
  • master-slave
  • robot
  • single port surgery

ASJC Scopus subject areas

  • Computer Science Applications
  • Biophysics
  • Surgery

Cite this

Kobayashi, Y., Tomono, Y., Sekiguchi, Y., Watanabe, H., Toyoda, K., Konishi, K., ... Fujie, M. G. (2010). A surgical robot with vision field control for single port endoscopic surgery. International Journal of Medical Robotics and Computer Assisted Surgery, 6(4), 454-464. https://doi.org/10.1002/rcs.355

A surgical robot with vision field control for single port endoscopic surgery. / Kobayashi, Yo; Tomono, Yu; Sekiguchi, Yuta; Watanabe, Hiroki; Toyoda, Kazutaka; Konishi, Kozo; Tomikawa, Morimasa; Ieiri, Satoshi; Tanoue, Kazuo; Hashizume, Makoto; Fujie, Masaktsu G.

In: International Journal of Medical Robotics and Computer Assisted Surgery, Vol. 6, No. 4, 12.2010, p. 454-464.

Research output: Contribution to journalArticle

Kobayashi, Y, Tomono, Y, Sekiguchi, Y, Watanabe, H, Toyoda, K, Konishi, K, Tomikawa, M, Ieiri, S, Tanoue, K, Hashizume, M & Fujie, MG 2010, 'A surgical robot with vision field control for single port endoscopic surgery', International Journal of Medical Robotics and Computer Assisted Surgery, vol. 6, no. 4, pp. 454-464. https://doi.org/10.1002/rcs.355
Kobayashi, Yo ; Tomono, Yu ; Sekiguchi, Yuta ; Watanabe, Hiroki ; Toyoda, Kazutaka ; Konishi, Kozo ; Tomikawa, Morimasa ; Ieiri, Satoshi ; Tanoue, Kazuo ; Hashizume, Makoto ; Fujie, Masaktsu G. / A surgical robot with vision field control for single port endoscopic surgery. In: International Journal of Medical Robotics and Computer Assisted Surgery. 2010 ; Vol. 6, No. 4. pp. 454-464.
@article{797c6dc89e5d411da1d68c5dc59ca168,
title = "A surgical robot with vision field control for single port endoscopic surgery",
abstract = "Background Robotic end-effectors for single port endoscopic surgery (SPS) require a manual change of vision field that slows surgery and increases the degrees of freedom (DOFs) of the manipulator. Methods A new surgical prototype robot has dynamic vision field control and a master controller to manipulate the endoscopic view. It uses positioning (4 DOF) and sheath (2 DOF) manipulators for vision field control, and dual tool tissue manipulators (gripping, 5 DOF; cautery, 3 DOF). Results The robot is feasible in vitro. 'Cut and vision field control' (using tool manipulators) was suitable for precise cutting tasks in risky areas; 'cut by vision field control' (using the vision field control manipulator) was effective for rapid macro cutting of tissues. A resection was performed using a combination of both methods. Conclusions The novel robotic system is feasible, but further studies are needed to address its performance in vivo.",
keywords = "endoscopic, master-slave, robot, single port surgery",
author = "Yo Kobayashi and Yu Tomono and Yuta Sekiguchi and Hiroki Watanabe and Kazutaka Toyoda and Kozo Konishi and Morimasa Tomikawa and Satoshi Ieiri and Kazuo Tanoue and Makoto Hashizume and Fujie, {Masaktsu G.}",
year = "2010",
month = "12",
doi = "10.1002/rcs.355",
language = "English",
volume = "6",
pages = "454--464",
journal = "International Journal of Medical Robotics and Computer Assisted Surgery",
issn = "1478-596X",
publisher = "John Wiley and Sons Ltd",
number = "4",

}

TY - JOUR

T1 - A surgical robot with vision field control for single port endoscopic surgery

AU - Kobayashi, Yo

AU - Tomono, Yu

AU - Sekiguchi, Yuta

AU - Watanabe, Hiroki

AU - Toyoda, Kazutaka

AU - Konishi, Kozo

AU - Tomikawa, Morimasa

AU - Ieiri, Satoshi

AU - Tanoue, Kazuo

AU - Hashizume, Makoto

AU - Fujie, Masaktsu G.

PY - 2010/12

Y1 - 2010/12

N2 - Background Robotic end-effectors for single port endoscopic surgery (SPS) require a manual change of vision field that slows surgery and increases the degrees of freedom (DOFs) of the manipulator. Methods A new surgical prototype robot has dynamic vision field control and a master controller to manipulate the endoscopic view. It uses positioning (4 DOF) and sheath (2 DOF) manipulators for vision field control, and dual tool tissue manipulators (gripping, 5 DOF; cautery, 3 DOF). Results The robot is feasible in vitro. 'Cut and vision field control' (using tool manipulators) was suitable for precise cutting tasks in risky areas; 'cut by vision field control' (using the vision field control manipulator) was effective for rapid macro cutting of tissues. A resection was performed using a combination of both methods. Conclusions The novel robotic system is feasible, but further studies are needed to address its performance in vivo.

AB - Background Robotic end-effectors for single port endoscopic surgery (SPS) require a manual change of vision field that slows surgery and increases the degrees of freedom (DOFs) of the manipulator. Methods A new surgical prototype robot has dynamic vision field control and a master controller to manipulate the endoscopic view. It uses positioning (4 DOF) and sheath (2 DOF) manipulators for vision field control, and dual tool tissue manipulators (gripping, 5 DOF; cautery, 3 DOF). Results The robot is feasible in vitro. 'Cut and vision field control' (using tool manipulators) was suitable for precise cutting tasks in risky areas; 'cut by vision field control' (using the vision field control manipulator) was effective for rapid macro cutting of tissues. A resection was performed using a combination of both methods. Conclusions The novel robotic system is feasible, but further studies are needed to address its performance in vivo.

KW - endoscopic

KW - master-slave

KW - robot

KW - single port surgery

UR - http://www.scopus.com/inward/record.url?scp=78649679734&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78649679734&partnerID=8YFLogxK

U2 - 10.1002/rcs.355

DO - 10.1002/rcs.355

M3 - Article

VL - 6

SP - 454

EP - 464

JO - International Journal of Medical Robotics and Computer Assisted Surgery

JF - International Journal of Medical Robotics and Computer Assisted Surgery

SN - 1478-596X

IS - 4

ER -