Abstract

An experimental system for the absolute measurements of He re-emission has been constructed. It mainly consists of a quadrupole mass spectrometer and specimen annealing equipments which connect to a 300 keV ion beam accelerator, and its operation can be controlled by a computer. Its constitution and performances, and examples of experimental results are reported. For the purpose of investigating the factors which inhibit blister formation or promote gas re-emission. He re-emission from single crystals and alloys has been measured during specimen annealing after irradiation with 100 keVe+ ion beam. By using an electron gun heater, a specimen can be annealed from room temperature to 2,000K at a stepwise rate of 50 K/min. Almost all of the implanted helium can be detected over the fluences of 2 X 1013 to 1 X 1017 ions/cm2.

Original languageEnglish
Pages (from-to)259-265
Number of pages7
JournalShinku/Journal of the Vacuum Society of Japan
Volume26
Issue number3
DOIs
Publication statusPublished - 1983
Externally publishedYes

Fingerprint

Helium
Ion beams
helium
Annealing
Electron guns
Mass spectrometers
ion beams
Particle accelerators
blisters
annealing
constitution
Gases
electron guns
Irradiation
Single crystals
Ions
heaters
mass spectrometers
fluence
accelerators

ASJC Scopus subject areas

  • Condensed Matter Physics
  • Surfaces, Coatings and Films
  • Electrical and Electronic Engineering

Cite this

Apparatus for Absolute Measurements of Helium Re-emission. / Nishijima, Toshiji; Kobayashi, Naoto.

In: Shinku/Journal of the Vacuum Society of Japan, Vol. 26, No. 3, 1983, p. 259-265.

Research output: Contribution to journalArticle

@article{746858a5eeb045328723eef87765a539,
title = "Apparatus for Absolute Measurements of Helium Re-emission",
abstract = "An experimental system for the absolute measurements of He re-emission has been constructed. It mainly consists of a quadrupole mass spectrometer and specimen annealing equipments which connect to a 300 keV ion beam accelerator, and its operation can be controlled by a computer. Its constitution and performances, and examples of experimental results are reported. For the purpose of investigating the factors which inhibit blister formation or promote gas re-emission. He re-emission from single crystals and alloys has been measured during specimen annealing after irradiation with 100 keVe+ ion beam. By using an electron gun heater, a specimen can be annealed from room temperature to 2,000K at a stepwise rate of 50 K/min. Almost all of the implanted helium can be detected over the fluences of 2 X 1013 to 1 X 1017 ions/cm2.",
author = "Toshiji Nishijima and Naoto Kobayashi",
year = "1983",
doi = "10.3131/jvsj.26.259",
language = "English",
volume = "26",
pages = "259--265",
journal = "Journal of the Vacuum Society of Japan",
issn = "1882-2398",
publisher = "Vacuum Society of Japan",
number = "3",

}

TY - JOUR

T1 - Apparatus for Absolute Measurements of Helium Re-emission

AU - Nishijima, Toshiji

AU - Kobayashi, Naoto

PY - 1983

Y1 - 1983

N2 - An experimental system for the absolute measurements of He re-emission has been constructed. It mainly consists of a quadrupole mass spectrometer and specimen annealing equipments which connect to a 300 keV ion beam accelerator, and its operation can be controlled by a computer. Its constitution and performances, and examples of experimental results are reported. For the purpose of investigating the factors which inhibit blister formation or promote gas re-emission. He re-emission from single crystals and alloys has been measured during specimen annealing after irradiation with 100 keVe+ ion beam. By using an electron gun heater, a specimen can be annealed from room temperature to 2,000K at a stepwise rate of 50 K/min. Almost all of the implanted helium can be detected over the fluences of 2 X 1013 to 1 X 1017 ions/cm2.

AB - An experimental system for the absolute measurements of He re-emission has been constructed. It mainly consists of a quadrupole mass spectrometer and specimen annealing equipments which connect to a 300 keV ion beam accelerator, and its operation can be controlled by a computer. Its constitution and performances, and examples of experimental results are reported. For the purpose of investigating the factors which inhibit blister formation or promote gas re-emission. He re-emission from single crystals and alloys has been measured during specimen annealing after irradiation with 100 keVe+ ion beam. By using an electron gun heater, a specimen can be annealed from room temperature to 2,000K at a stepwise rate of 50 K/min. Almost all of the implanted helium can be detected over the fluences of 2 X 1013 to 1 X 1017 ions/cm2.

UR - http://www.scopus.com/inward/record.url?scp=85004575621&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85004575621&partnerID=8YFLogxK

U2 - 10.3131/jvsj.26.259

DO - 10.3131/jvsj.26.259

M3 - Article

AN - SCOPUS:85004575621

VL - 26

SP - 259

EP - 265

JO - Journal of the Vacuum Society of Japan

JF - Journal of the Vacuum Society of Japan

SN - 1882-2398

IS - 3

ER -