Construction of a high-resolution electron energy loss spectrometer

Tadaaki Nagao, Yoshitaka Iizuka, Makoto Umeuchi, Teruhiko Shimazaki, Michio Nakajima, Chuhei Oshima

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

A new version of a high-resolution electron energy loss spectrometer has been constructed. Major modifications are the enlargement of the analyzer, gold coating of all the electrode surfaces facing the electron beam, and the reduction of magnetic fields. A high detector current of 4 pA has been obtained together with the narrow energy half width of 1.2 meV (FWHM) in the straight-through mode.

Original languageEnglish
Pages (from-to)515-516
Number of pages2
JournalReview of Scientific Instruments
Volume65
Issue number2
DOIs
Publication statusPublished - 1994

Fingerprint

Gold coatings
gold coatings
Full width at half maximum
Spectrometers
Electron beams
analyzers
Energy dissipation
energy dissipation
electron beams
spectrometers
electron energy
Magnetic fields
Detectors
Electrodes
electrodes
Electrons
high resolution
detectors
magnetic fields
energy

ASJC Scopus subject areas

  • Instrumentation
  • Physics and Astronomy (miscellaneous)

Cite this

Nagao, T., Iizuka, Y., Umeuchi, M., Shimazaki, T., Nakajima, M., & Oshima, C. (1994). Construction of a high-resolution electron energy loss spectrometer. Review of Scientific Instruments, 65(2), 515-516. https://doi.org/10.1063/1.1145115

Construction of a high-resolution electron energy loss spectrometer. / Nagao, Tadaaki; Iizuka, Yoshitaka; Umeuchi, Makoto; Shimazaki, Teruhiko; Nakajima, Michio; Oshima, Chuhei.

In: Review of Scientific Instruments, Vol. 65, No. 2, 1994, p. 515-516.

Research output: Contribution to journalArticle

Nagao, T, Iizuka, Y, Umeuchi, M, Shimazaki, T, Nakajima, M & Oshima, C 1994, 'Construction of a high-resolution electron energy loss spectrometer', Review of Scientific Instruments, vol. 65, no. 2, pp. 515-516. https://doi.org/10.1063/1.1145115
Nagao T, Iizuka Y, Umeuchi M, Shimazaki T, Nakajima M, Oshima C. Construction of a high-resolution electron energy loss spectrometer. Review of Scientific Instruments. 1994;65(2):515-516. https://doi.org/10.1063/1.1145115
Nagao, Tadaaki ; Iizuka, Yoshitaka ; Umeuchi, Makoto ; Shimazaki, Teruhiko ; Nakajima, Michio ; Oshima, Chuhei. / Construction of a high-resolution electron energy loss spectrometer. In: Review of Scientific Instruments. 1994 ; Vol. 65, No. 2. pp. 515-516.
@article{ab7da36b2df74786a9d07da59c9b0c91,
title = "Construction of a high-resolution electron energy loss spectrometer",
abstract = "A new version of a high-resolution electron energy loss spectrometer has been constructed. Major modifications are the enlargement of the analyzer, gold coating of all the electrode surfaces facing the electron beam, and the reduction of magnetic fields. A high detector current of 4 pA has been obtained together with the narrow energy half width of 1.2 meV (FWHM) in the straight-through mode.",
author = "Tadaaki Nagao and Yoshitaka Iizuka and Makoto Umeuchi and Teruhiko Shimazaki and Michio Nakajima and Chuhei Oshima",
year = "1994",
doi = "10.1063/1.1145115",
language = "English",
volume = "65",
pages = "515--516",
journal = "Review of Scientific Instruments",
issn = "0034-6748",
publisher = "American Institute of Physics Publising LLC",
number = "2",

}

TY - JOUR

T1 - Construction of a high-resolution electron energy loss spectrometer

AU - Nagao, Tadaaki

AU - Iizuka, Yoshitaka

AU - Umeuchi, Makoto

AU - Shimazaki, Teruhiko

AU - Nakajima, Michio

AU - Oshima, Chuhei

PY - 1994

Y1 - 1994

N2 - A new version of a high-resolution electron energy loss spectrometer has been constructed. Major modifications are the enlargement of the analyzer, gold coating of all the electrode surfaces facing the electron beam, and the reduction of magnetic fields. A high detector current of 4 pA has been obtained together with the narrow energy half width of 1.2 meV (FWHM) in the straight-through mode.

AB - A new version of a high-resolution electron energy loss spectrometer has been constructed. Major modifications are the enlargement of the analyzer, gold coating of all the electrode surfaces facing the electron beam, and the reduction of magnetic fields. A high detector current of 4 pA has been obtained together with the narrow energy half width of 1.2 meV (FWHM) in the straight-through mode.

UR - http://www.scopus.com/inward/record.url?scp=36448998667&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=36448998667&partnerID=8YFLogxK

U2 - 10.1063/1.1145115

DO - 10.1063/1.1145115

M3 - Article

VL - 65

SP - 515

EP - 516

JO - Review of Scientific Instruments

JF - Review of Scientific Instruments

SN - 0034-6748

IS - 2

ER -