Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery.

K. Masamune, E. Kobayashi, Y. Masutani, M. Suzuki, T. Dohi, Hiroshi Iseki, K. Takakura

Research output: Contribution to journalArticle

207 Citations (Scopus)

Abstract

A variety of medical robots for stereotactic neurosurgery has been developed in recent years. Almost of all these robots use computed tomography (CT) to scan the brain of the patient before and during surgery. Currently, we are developing a needle insertion manipulator for magnetic resonance imaging (MRI)-guided neurosurgery. MRI techniques, including MRI angiography and functional MRI, are attractive for the development of interventional MRI therapies and operations. If a robot were available, these therapies would be minimally invasive, with more accurate guidance than is possible with current CT-guided systems. Actuation of a robot in an MRI environment is difficult because of the presence of strong magnetic fields. Therefore, the robot must be constructed of nonmagnetic materials. The system frame was manufactured using polyethylene terephthalate (PET) and was actuated using ultrasonic motors. Accuracy-evaluation procedures and phantom tests have been performed. The total accuracy of the system was approximately 3.0 mm. No artifacts caused by the manipulator were observed in the images.

Original languageEnglish
Pages (from-to)242-248
Number of pages7
JournalJournal of image guided surgery
Volume1
Issue number4
Publication statusPublished - 1995
Externally publishedYes

Fingerprint

Neurosurgery
Magnetic resonance
Needles
Manipulators
Magnetic Resonance Imaging
Robots
Imaging techniques
Tomography
Interventional Magnetic Resonance Imaging
Polyethylene Terephthalates
Angiography
Magnetic Resonance Angiography
Magnetic Fields
Ultrasonics
Polyethylene terephthalates
Artifacts
Surgery
Brain
Magnetic fields
Therapeutics

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Medicine (miscellaneous)
  • Surgery

Cite this

Masamune, K., Kobayashi, E., Masutani, Y., Suzuki, M., Dohi, T., Iseki, H., & Takakura, K. (1995). Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. Journal of image guided surgery, 1(4), 242-248.

Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. / Masamune, K.; Kobayashi, E.; Masutani, Y.; Suzuki, M.; Dohi, T.; Iseki, Hiroshi; Takakura, K.

In: Journal of image guided surgery, Vol. 1, No. 4, 1995, p. 242-248.

Research output: Contribution to journalArticle

Masamune, K, Kobayashi, E, Masutani, Y, Suzuki, M, Dohi, T, Iseki, H & Takakura, K 1995, 'Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery.', Journal of image guided surgery, vol. 1, no. 4, pp. 242-248.
Masamune K, Kobayashi E, Masutani Y, Suzuki M, Dohi T, Iseki H et al. Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. Journal of image guided surgery. 1995;1(4):242-248.
Masamune, K. ; Kobayashi, E. ; Masutani, Y. ; Suzuki, M. ; Dohi, T. ; Iseki, Hiroshi ; Takakura, K. / Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. In: Journal of image guided surgery. 1995 ; Vol. 1, No. 4. pp. 242-248.
@article{04319cb3a7704876a9b8601add5e4450,
title = "Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery.",
abstract = "A variety of medical robots for stereotactic neurosurgery has been developed in recent years. Almost of all these robots use computed tomography (CT) to scan the brain of the patient before and during surgery. Currently, we are developing a needle insertion manipulator for magnetic resonance imaging (MRI)-guided neurosurgery. MRI techniques, including MRI angiography and functional MRI, are attractive for the development of interventional MRI therapies and operations. If a robot were available, these therapies would be minimally invasive, with more accurate guidance than is possible with current CT-guided systems. Actuation of a robot in an MRI environment is difficult because of the presence of strong magnetic fields. Therefore, the robot must be constructed of nonmagnetic materials. The system frame was manufactured using polyethylene terephthalate (PET) and was actuated using ultrasonic motors. Accuracy-evaluation procedures and phantom tests have been performed. The total accuracy of the system was approximately 3.0 mm. No artifacts caused by the manipulator were observed in the images.",
author = "K. Masamune and E. Kobayashi and Y. Masutani and M. Suzuki and T. Dohi and Hiroshi Iseki and K. Takakura",
year = "1995",
language = "English",
volume = "1",
pages = "242--248",
journal = "Computer Assisted Surgery",
issn = "1092-9088",
publisher = "Taylor and Francis Ltd.",
number = "4",

}

TY - JOUR

T1 - Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery.

AU - Masamune, K.

AU - Kobayashi, E.

AU - Masutani, Y.

AU - Suzuki, M.

AU - Dohi, T.

AU - Iseki, Hiroshi

AU - Takakura, K.

PY - 1995

Y1 - 1995

N2 - A variety of medical robots for stereotactic neurosurgery has been developed in recent years. Almost of all these robots use computed tomography (CT) to scan the brain of the patient before and during surgery. Currently, we are developing a needle insertion manipulator for magnetic resonance imaging (MRI)-guided neurosurgery. MRI techniques, including MRI angiography and functional MRI, are attractive for the development of interventional MRI therapies and operations. If a robot were available, these therapies would be minimally invasive, with more accurate guidance than is possible with current CT-guided systems. Actuation of a robot in an MRI environment is difficult because of the presence of strong magnetic fields. Therefore, the robot must be constructed of nonmagnetic materials. The system frame was manufactured using polyethylene terephthalate (PET) and was actuated using ultrasonic motors. Accuracy-evaluation procedures and phantom tests have been performed. The total accuracy of the system was approximately 3.0 mm. No artifacts caused by the manipulator were observed in the images.

AB - A variety of medical robots for stereotactic neurosurgery has been developed in recent years. Almost of all these robots use computed tomography (CT) to scan the brain of the patient before and during surgery. Currently, we are developing a needle insertion manipulator for magnetic resonance imaging (MRI)-guided neurosurgery. MRI techniques, including MRI angiography and functional MRI, are attractive for the development of interventional MRI therapies and operations. If a robot were available, these therapies would be minimally invasive, with more accurate guidance than is possible with current CT-guided systems. Actuation of a robot in an MRI environment is difficult because of the presence of strong magnetic fields. Therefore, the robot must be constructed of nonmagnetic materials. The system frame was manufactured using polyethylene terephthalate (PET) and was actuated using ultrasonic motors. Accuracy-evaluation procedures and phantom tests have been performed. The total accuracy of the system was approximately 3.0 mm. No artifacts caused by the manipulator were observed in the images.

UR - http://www.scopus.com/inward/record.url?scp=0029422530&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029422530&partnerID=8YFLogxK

M3 - Article

VL - 1

SP - 242

EP - 248

JO - Computer Assisted Surgery

JF - Computer Assisted Surgery

SN - 1092-9088

IS - 4

ER -