Development of in-vitro evaluation system for annuloplasty rings

M. Arita, H. Kasegawa, M. Umezu

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

An in-vitro system was devised to study the performance of newly developed annuloplasty rings as well as various conventional rings. The Duran Flexible Annuloplasty Ring was tested to define its characteristics and to validate the system for comparative testing. It was possible to obtain quantitative data using a microscope and a load cell to respectively measure valve orifice area and tensile load on the valve annulus. The results suggest that this apparatus could be employed to characterize the features and functional performance of other types of annuloplasty ring.

Original languageEnglish
Pages (from-to)14-18
Number of pages5
JournalAsian Cardiovascular and Thoracic Annals
Volume9
Issue number1
DOIs
Publication statusPublished - 2001 Jan 1

    Fingerprint

ASJC Scopus subject areas

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine