Development of simple proton CT system with novel correction methods of proton scattering

M. Takabe, T. Masuda, M. Arimoto, Jun Kataoka, K. Sueoka, T. Maruhashi, S. Tanaka, T. Nishio, T. Toshito, M. Kimura, T. Inaniwa

    Research output: Contribution to journalArticle

    Abstract

    For safe and effective proton therapy, the proton range in a patient's body is characterized by the water equivalent length (WEL), and must be accurately determined. Current treatment planning is based on X-ray computed tomography images, which might cause uncertainty because of the different energy loss processes between protons and X-rays. We develop a simple, novel, and real-time proton CT system. The system uses a CCD camera and scintillator, which is thin enough for protons to penetrate. Since protons lose energy when they pass through a phantom, different emissions corresponding to the proton energy loss are acquired in the scintillator. Images of the scintillator were gathered by the CCD camera with 70 MeV and 200 MeV proton beams. Since blurring due to proton reactions such as multiple Coulomb scattering and nuclear reactions significantly degrades the obtained images in both beams, we developed two kinds of effective correction methods. One method is applied to broad beam systems, while the other is applied to narrow beam systems. We successfully obtain clear images with minor proton reaction effects by applying these correction methods. Moreover, we confirm that the WEL values estimated from the acquired CT images agree well with the theoretical values for materials such as polymethyl methacrylate (PMMA) and isopropyl alcohol, within 1-σ uncertainty. Through simulations, we found that nuclear reactions significantly contribute to the uncertainty of WEL values.

    Fingerprint

    proton scattering
    Protons
    Scattering
    protons
    scintillation counters
    Phosphors
    Nuclear reactions
    CCD cameras
    nuclear reactions
    energy dissipation
    Energy dissipation
    water
    isopropyl alcohol
    blurring
    Water
    X rays
    Proton beams
    proton energy
    proton beams
    polymethyl methacrylate

    Keywords

    • CCD camera
    • Multiple Coulomb scattering (MCS)
    • Proton computed tomography (pCT)
    • Proton therapy
    • Range measurement
    • Water equivalent length (WEL)

    ASJC Scopus subject areas

    • Nuclear and High Energy Physics
    • Instrumentation

    Cite this

    Development of simple proton CT system with novel correction methods of proton scattering. / Takabe, M.; Masuda, T.; Arimoto, M.; Kataoka, Jun; Sueoka, K.; Maruhashi, T.; Tanaka, S.; Nishio, T.; Toshito, T.; Kimura, M.; Inaniwa, T.

    In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 01.01.2018.

    Research output: Contribution to journalArticle

    @article{e2311f83a23d4bceb7a1344ab4c46ee6,
    title = "Development of simple proton CT system with novel correction methods of proton scattering",
    abstract = "For safe and effective proton therapy, the proton range in a patient's body is characterized by the water equivalent length (WEL), and must be accurately determined. Current treatment planning is based on X-ray computed tomography images, which might cause uncertainty because of the different energy loss processes between protons and X-rays. We develop a simple, novel, and real-time proton CT system. The system uses a CCD camera and scintillator, which is thin enough for protons to penetrate. Since protons lose energy when they pass through a phantom, different emissions corresponding to the proton energy loss are acquired in the scintillator. Images of the scintillator were gathered by the CCD camera with 70 MeV and 200 MeV proton beams. Since blurring due to proton reactions such as multiple Coulomb scattering and nuclear reactions significantly degrades the obtained images in both beams, we developed two kinds of effective correction methods. One method is applied to broad beam systems, while the other is applied to narrow beam systems. We successfully obtain clear images with minor proton reaction effects by applying these correction methods. Moreover, we confirm that the WEL values estimated from the acquired CT images agree well with the theoretical values for materials such as polymethyl methacrylate (PMMA) and isopropyl alcohol, within 1-σ uncertainty. Through simulations, we found that nuclear reactions significantly contribute to the uncertainty of WEL values.",
    keywords = "CCD camera, Multiple Coulomb scattering (MCS), Proton computed tomography (pCT), Proton therapy, Range measurement, Water equivalent length (WEL)",
    author = "M. Takabe and T. Masuda and M. Arimoto and Jun Kataoka and K. Sueoka and T. Maruhashi and S. Tanaka and T. Nishio and T. Toshito and M. Kimura and T. Inaniwa",
    year = "2018",
    month = "1",
    day = "1",
    doi = "10.1016/j.nima.2018.05.034",
    language = "English",
    journal = "Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment",
    issn = "0168-9002",
    publisher = "Elsevier",

    }

    TY - JOUR

    T1 - Development of simple proton CT system with novel correction methods of proton scattering

    AU - Takabe, M.

    AU - Masuda, T.

    AU - Arimoto, M.

    AU - Kataoka, Jun

    AU - Sueoka, K.

    AU - Maruhashi, T.

    AU - Tanaka, S.

    AU - Nishio, T.

    AU - Toshito, T.

    AU - Kimura, M.

    AU - Inaniwa, T.

    PY - 2018/1/1

    Y1 - 2018/1/1

    N2 - For safe and effective proton therapy, the proton range in a patient's body is characterized by the water equivalent length (WEL), and must be accurately determined. Current treatment planning is based on X-ray computed tomography images, which might cause uncertainty because of the different energy loss processes between protons and X-rays. We develop a simple, novel, and real-time proton CT system. The system uses a CCD camera and scintillator, which is thin enough for protons to penetrate. Since protons lose energy when they pass through a phantom, different emissions corresponding to the proton energy loss are acquired in the scintillator. Images of the scintillator were gathered by the CCD camera with 70 MeV and 200 MeV proton beams. Since blurring due to proton reactions such as multiple Coulomb scattering and nuclear reactions significantly degrades the obtained images in both beams, we developed two kinds of effective correction methods. One method is applied to broad beam systems, while the other is applied to narrow beam systems. We successfully obtain clear images with minor proton reaction effects by applying these correction methods. Moreover, we confirm that the WEL values estimated from the acquired CT images agree well with the theoretical values for materials such as polymethyl methacrylate (PMMA) and isopropyl alcohol, within 1-σ uncertainty. Through simulations, we found that nuclear reactions significantly contribute to the uncertainty of WEL values.

    AB - For safe and effective proton therapy, the proton range in a patient's body is characterized by the water equivalent length (WEL), and must be accurately determined. Current treatment planning is based on X-ray computed tomography images, which might cause uncertainty because of the different energy loss processes between protons and X-rays. We develop a simple, novel, and real-time proton CT system. The system uses a CCD camera and scintillator, which is thin enough for protons to penetrate. Since protons lose energy when they pass through a phantom, different emissions corresponding to the proton energy loss are acquired in the scintillator. Images of the scintillator were gathered by the CCD camera with 70 MeV and 200 MeV proton beams. Since blurring due to proton reactions such as multiple Coulomb scattering and nuclear reactions significantly degrades the obtained images in both beams, we developed two kinds of effective correction methods. One method is applied to broad beam systems, while the other is applied to narrow beam systems. We successfully obtain clear images with minor proton reaction effects by applying these correction methods. Moreover, we confirm that the WEL values estimated from the acquired CT images agree well with the theoretical values for materials such as polymethyl methacrylate (PMMA) and isopropyl alcohol, within 1-σ uncertainty. Through simulations, we found that nuclear reactions significantly contribute to the uncertainty of WEL values.

    KW - CCD camera

    KW - Multiple Coulomb scattering (MCS)

    KW - Proton computed tomography (pCT)

    KW - Proton therapy

    KW - Range measurement

    KW - Water equivalent length (WEL)

    UR - http://www.scopus.com/inward/record.url?scp=85050867964&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=85050867964&partnerID=8YFLogxK

    U2 - 10.1016/j.nima.2018.05.034

    DO - 10.1016/j.nima.2018.05.034

    M3 - Article

    JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

    JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

    SN - 0168-9002

    ER -