Development of the multiple sample injector using stepwise flow to prevent sample dilution

Masaki Kanai, Shingo Ikeda, Junichi Tanaka, Jeung Sang Go, Hiroaki Nakanishi, Shuichi Shoji

    Research output: Chapter in Book/Report/Conference proceedingConference contribution

    1 Citation (Scopus)

    Abstract

    This paper presents a concentration-conservative multiple sample injector. The stepwise flow was used to reduce dilution of the sample reagent while the reagent was injected into the sample collection outlet. By using CFD simulation, the concentration profile of the sample reagent was estimated. It was clearly shown the concentration was kept in a sample outlet. For experimental evaluation, a sample injector was fabricated and the sample injection using stepwise flow was observed by fluorescent measuring system.

    Original languageEnglish
    Title of host publicationProceedings of the IEEE Micro Electro Mechanical Systems (MEMS)
    Pages189-192
    Number of pages4
    Publication statusPublished - 2003
    EventIEEE Sixteenth Annual International Conference on Micro Electro Mechanical Systems - Kyoto
    Duration: 2003 Jan 192003 Jan 23

    Other

    OtherIEEE Sixteenth Annual International Conference on Micro Electro Mechanical Systems
    CityKyoto
    Period03/1/1903/1/23

    Fingerprint

    Dilution
    Computational fluid dynamics

    ASJC Scopus subject areas

    • Mechanical Engineering
    • Electrical and Electronic Engineering
    • Control and Systems Engineering

    Cite this

    Kanai, M., Ikeda, S., Tanaka, J., Go, J. S., Nakanishi, H., & Shoji, S. (2003). Development of the multiple sample injector using stepwise flow to prevent sample dilution. In Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS) (pp. 189-192)

    Development of the multiple sample injector using stepwise flow to prevent sample dilution. / Kanai, Masaki; Ikeda, Shingo; Tanaka, Junichi; Go, Jeung Sang; Nakanishi, Hiroaki; Shoji, Shuichi.

    Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS). 2003. p. 189-192.

    Research output: Chapter in Book/Report/Conference proceedingConference contribution

    Kanai, M, Ikeda, S, Tanaka, J, Go, JS, Nakanishi, H & Shoji, S 2003, Development of the multiple sample injector using stepwise flow to prevent sample dilution. in Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS). pp. 189-192, IEEE Sixteenth Annual International Conference on Micro Electro Mechanical Systems, Kyoto, 03/1/19.
    Kanai M, Ikeda S, Tanaka J, Go JS, Nakanishi H, Shoji S. Development of the multiple sample injector using stepwise flow to prevent sample dilution. In Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS). 2003. p. 189-192
    Kanai, Masaki ; Ikeda, Shingo ; Tanaka, Junichi ; Go, Jeung Sang ; Nakanishi, Hiroaki ; Shoji, Shuichi. / Development of the multiple sample injector using stepwise flow to prevent sample dilution. Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS). 2003. pp. 189-192
    @inproceedings{eb6345d4cc7a4a8c93d8acdf632108ce,
    title = "Development of the multiple sample injector using stepwise flow to prevent sample dilution",
    abstract = "This paper presents a concentration-conservative multiple sample injector. The stepwise flow was used to reduce dilution of the sample reagent while the reagent was injected into the sample collection outlet. By using CFD simulation, the concentration profile of the sample reagent was estimated. It was clearly shown the concentration was kept in a sample outlet. For experimental evaluation, a sample injector was fabricated and the sample injection using stepwise flow was observed by fluorescent measuring system.",
    author = "Masaki Kanai and Shingo Ikeda and Junichi Tanaka and Go, {Jeung Sang} and Hiroaki Nakanishi and Shuichi Shoji",
    year = "2003",
    language = "English",
    pages = "189--192",
    booktitle = "Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS)",

    }

    TY - GEN

    T1 - Development of the multiple sample injector using stepwise flow to prevent sample dilution

    AU - Kanai, Masaki

    AU - Ikeda, Shingo

    AU - Tanaka, Junichi

    AU - Go, Jeung Sang

    AU - Nakanishi, Hiroaki

    AU - Shoji, Shuichi

    PY - 2003

    Y1 - 2003

    N2 - This paper presents a concentration-conservative multiple sample injector. The stepwise flow was used to reduce dilution of the sample reagent while the reagent was injected into the sample collection outlet. By using CFD simulation, the concentration profile of the sample reagent was estimated. It was clearly shown the concentration was kept in a sample outlet. For experimental evaluation, a sample injector was fabricated and the sample injection using stepwise flow was observed by fluorescent measuring system.

    AB - This paper presents a concentration-conservative multiple sample injector. The stepwise flow was used to reduce dilution of the sample reagent while the reagent was injected into the sample collection outlet. By using CFD simulation, the concentration profile of the sample reagent was estimated. It was clearly shown the concentration was kept in a sample outlet. For experimental evaluation, a sample injector was fabricated and the sample injection using stepwise flow was observed by fluorescent measuring system.

    UR - http://www.scopus.com/inward/record.url?scp=0038156088&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=0038156088&partnerID=8YFLogxK

    M3 - Conference contribution

    AN - SCOPUS:0038156088

    SP - 189

    EP - 192

    BT - Proceedings of the IEEE Micro Electro Mechanical Systems (MEMS)

    ER -