Load-dispersing design with twined-spring geometry of a distensible intracranial stent for cerebral aneurysms

Yasuhiro Shobayashi, Satoshi Tateshima, Kazuo Tanishita

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Endovascular stents are increasingly being used to treat cerebral aneurysms. Mechanically, a cerebrovascular stent must have a low radial stiffness to prevent vessel dissection and rupture. To minimize these complications, we need to consider a stent design that has a low radial force and disperses the load within the stented artery. Therefore, highly distensible, load-dispersion stent designs are desirable for intracranial stenting. This study focused on closed-cell stent geometries and calculated the differences in stress within the artery because of the structure by using finite-element modeling. The results showed that the design with hexagonal cell geometry stretched in the circumferential direction had lower radial and circumferential stresses than did the other models. Comparing the maximum radial stress of our models, stress reduction of 35% was obtained with this design. Moreover, its radial stress was 47 kPa, which was similar to the critical stress of 42 kPa assumed in this study. This stent model was characterized by narrow strut spacing and a large surface area, which was dominated by the twined-spring geometry. It had low radial and circumferential stresses and a dispersed stress distribution compared with the other models. Therefore, this design is a desirable load-dispersing design for cerebrovascular treatment.

Original languageEnglish
Pages (from-to)29-37
Number of pages9
JournalJournal of Biorheology
Volume26
Issue number1-2
DOIs
Publication statusPublished - 2013
Externally publishedYes

Fingerprint

Stents
Geometry
Dissection
Struts
Stress concentration
Stiffness

Keywords

  • Cerebral aneurysm
  • Flexibility
  • Radial stiffness
  • Stent
  • Stent-vessel interactions

ASJC Scopus subject areas

  • Mechanics of Materials
  • Materials Science(all)
  • Mechanical Engineering

Cite this

Load-dispersing design with twined-spring geometry of a distensible intracranial stent for cerebral aneurysms. / Shobayashi, Yasuhiro; Tateshima, Satoshi; Tanishita, Kazuo.

In: Journal of Biorheology, Vol. 26, No. 1-2, 2013, p. 29-37.

Research output: Contribution to journalArticle

Shobayashi, Yasuhiro ; Tateshima, Satoshi ; Tanishita, Kazuo. / Load-dispersing design with twined-spring geometry of a distensible intracranial stent for cerebral aneurysms. In: Journal of Biorheology. 2013 ; Vol. 26, No. 1-2. pp. 29-37.
@article{8c5ddad64ec04daa96e1293d65544b51,
title = "Load-dispersing design with twined-spring geometry of a distensible intracranial stent for cerebral aneurysms",
abstract = "Endovascular stents are increasingly being used to treat cerebral aneurysms. Mechanically, a cerebrovascular stent must have a low radial stiffness to prevent vessel dissection and rupture. To minimize these complications, we need to consider a stent design that has a low radial force and disperses the load within the stented artery. Therefore, highly distensible, load-dispersion stent designs are desirable for intracranial stenting. This study focused on closed-cell stent geometries and calculated the differences in stress within the artery because of the structure by using finite-element modeling. The results showed that the design with hexagonal cell geometry stretched in the circumferential direction had lower radial and circumferential stresses than did the other models. Comparing the maximum radial stress of our models, stress reduction of 35{\%} was obtained with this design. Moreover, its radial stress was 47 kPa, which was similar to the critical stress of 42 kPa assumed in this study. This stent model was characterized by narrow strut spacing and a large surface area, which was dominated by the twined-spring geometry. It had low radial and circumferential stresses and a dispersed stress distribution compared with the other models. Therefore, this design is a desirable load-dispersing design for cerebrovascular treatment.",
keywords = "Cerebral aneurysm, Flexibility, Radial stiffness, Stent, Stent-vessel interactions",
author = "Yasuhiro Shobayashi and Satoshi Tateshima and Kazuo Tanishita",
year = "2013",
doi = "10.1007/s12573-011-0044-1",
language = "English",
volume = "26",
pages = "29--37",
journal = "Journal of Biorheology",
issn = "1867-0466",
publisher = "Springer Japan",
number = "1-2",

}

TY - JOUR

T1 - Load-dispersing design with twined-spring geometry of a distensible intracranial stent for cerebral aneurysms

AU - Shobayashi, Yasuhiro

AU - Tateshima, Satoshi

AU - Tanishita, Kazuo

PY - 2013

Y1 - 2013

N2 - Endovascular stents are increasingly being used to treat cerebral aneurysms. Mechanically, a cerebrovascular stent must have a low radial stiffness to prevent vessel dissection and rupture. To minimize these complications, we need to consider a stent design that has a low radial force and disperses the load within the stented artery. Therefore, highly distensible, load-dispersion stent designs are desirable for intracranial stenting. This study focused on closed-cell stent geometries and calculated the differences in stress within the artery because of the structure by using finite-element modeling. The results showed that the design with hexagonal cell geometry stretched in the circumferential direction had lower radial and circumferential stresses than did the other models. Comparing the maximum radial stress of our models, stress reduction of 35% was obtained with this design. Moreover, its radial stress was 47 kPa, which was similar to the critical stress of 42 kPa assumed in this study. This stent model was characterized by narrow strut spacing and a large surface area, which was dominated by the twined-spring geometry. It had low radial and circumferential stresses and a dispersed stress distribution compared with the other models. Therefore, this design is a desirable load-dispersing design for cerebrovascular treatment.

AB - Endovascular stents are increasingly being used to treat cerebral aneurysms. Mechanically, a cerebrovascular stent must have a low radial stiffness to prevent vessel dissection and rupture. To minimize these complications, we need to consider a stent design that has a low radial force and disperses the load within the stented artery. Therefore, highly distensible, load-dispersion stent designs are desirable for intracranial stenting. This study focused on closed-cell stent geometries and calculated the differences in stress within the artery because of the structure by using finite-element modeling. The results showed that the design with hexagonal cell geometry stretched in the circumferential direction had lower radial and circumferential stresses than did the other models. Comparing the maximum radial stress of our models, stress reduction of 35% was obtained with this design. Moreover, its radial stress was 47 kPa, which was similar to the critical stress of 42 kPa assumed in this study. This stent model was characterized by narrow strut spacing and a large surface area, which was dominated by the twined-spring geometry. It had low radial and circumferential stresses and a dispersed stress distribution compared with the other models. Therefore, this design is a desirable load-dispersing design for cerebrovascular treatment.

KW - Cerebral aneurysm

KW - Flexibility

KW - Radial stiffness

KW - Stent

KW - Stent-vessel interactions

UR - http://www.scopus.com/inward/record.url?scp=84875373151&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84875373151&partnerID=8YFLogxK

U2 - 10.1007/s12573-011-0044-1

DO - 10.1007/s12573-011-0044-1

M3 - Article

VL - 26

SP - 29

EP - 37

JO - Journal of Biorheology

JF - Journal of Biorheology

SN - 1867-0466

IS - 1-2

ER -