Mechanical-Based Model for Extra-Fine Needle Tip Deflection Until Breaching of Tissue Surface

    Research output: Contribution to journalArticle

    Abstract

    Accurate estimation of needle deflection is necessary to successfully steer the needle to targets located deep inside the body. In particular, the deflection that occurs until the tissue surface is breached differs according to the tissue shape and stiffness. This topic has not been a focus of previous work. In the present paper, we propose a model with which to estimate the needle deflection that occurs until breaching of the tissue surface with consideration of the tissue shape and stiffness. This model comprises a cantilever beam supported by virtual springs that represent the interaction forces between the needle tip and tissue surface. The effects of different insertion angles and tissue stiffness on needle deflection are represented by changing the spring constants. The model was used in experiments involving four different insertion angles (0º, 15º, 30º, and 45º) and three different polyvinyl chloride (PVC) phantoms with different stiffness (100, 75, and 50%). We verified the proposed model with the 80% PVC phantom and showed a maximum error of 0.04 mm.

    Original languageEnglish
    Pages (from-to)697-706
    Number of pages10
    JournalJournal of Medical and Biological Engineering
    Volume38
    Issue number5
    DOIs
    Publication statusPublished - 2018 Oct 1

    Fingerprint

    Needles
    Tissue
    Stiffness
    Polyvinyl chlorides
    Cantilever beams
    Experiments

    Keywords

    • Deflection estimation
    • Mechanical-based model
    • Needle steering
    • Rayleigh–Ritz method

    ASJC Scopus subject areas

    • Biomedical Engineering

    Cite this

    Mechanical-Based Model for Extra-Fine Needle Tip Deflection Until Breaching of Tissue Surface. / Tsumura, Ryosuke; Miyashita, Tomoyuki; Iwata, Hiroyasu.

    In: Journal of Medical and Biological Engineering, Vol. 38, No. 5, 01.10.2018, p. 697-706.

    Research output: Contribution to journalArticle

    @article{eae09493adfc4db3afdc6b9a02b49c0c,
    title = "Mechanical-Based Model for Extra-Fine Needle Tip Deflection Until Breaching of Tissue Surface",
    abstract = "Accurate estimation of needle deflection is necessary to successfully steer the needle to targets located deep inside the body. In particular, the deflection that occurs until the tissue surface is breached differs according to the tissue shape and stiffness. This topic has not been a focus of previous work. In the present paper, we propose a model with which to estimate the needle deflection that occurs until breaching of the tissue surface with consideration of the tissue shape and stiffness. This model comprises a cantilever beam supported by virtual springs that represent the interaction forces between the needle tip and tissue surface. The effects of different insertion angles and tissue stiffness on needle deflection are represented by changing the spring constants. The model was used in experiments involving four different insertion angles (0º, 15º, 30º, and 45º) and three different polyvinyl chloride (PVC) phantoms with different stiffness (100, 75, and 50{\%}). We verified the proposed model with the 80{\%} PVC phantom and showed a maximum error of 0.04 mm.",
    keywords = "Deflection estimation, Mechanical-based model, Needle steering, Rayleigh–Ritz method",
    author = "Ryosuke Tsumura and Tomoyuki Miyashita and Hiroyasu Iwata",
    year = "2018",
    month = "10",
    day = "1",
    doi = "10.1007/s40846-017-0359-5",
    language = "English",
    volume = "38",
    pages = "697--706",
    journal = "Journal of Medical and Biological Engineering",
    issn = "1609-0985",
    publisher = "Biomedical Engineering Society",
    number = "5",

    }

    TY - JOUR

    T1 - Mechanical-Based Model for Extra-Fine Needle Tip Deflection Until Breaching of Tissue Surface

    AU - Tsumura, Ryosuke

    AU - Miyashita, Tomoyuki

    AU - Iwata, Hiroyasu

    PY - 2018/10/1

    Y1 - 2018/10/1

    N2 - Accurate estimation of needle deflection is necessary to successfully steer the needle to targets located deep inside the body. In particular, the deflection that occurs until the tissue surface is breached differs according to the tissue shape and stiffness. This topic has not been a focus of previous work. In the present paper, we propose a model with which to estimate the needle deflection that occurs until breaching of the tissue surface with consideration of the tissue shape and stiffness. This model comprises a cantilever beam supported by virtual springs that represent the interaction forces between the needle tip and tissue surface. The effects of different insertion angles and tissue stiffness on needle deflection are represented by changing the spring constants. The model was used in experiments involving four different insertion angles (0º, 15º, 30º, and 45º) and three different polyvinyl chloride (PVC) phantoms with different stiffness (100, 75, and 50%). We verified the proposed model with the 80% PVC phantom and showed a maximum error of 0.04 mm.

    AB - Accurate estimation of needle deflection is necessary to successfully steer the needle to targets located deep inside the body. In particular, the deflection that occurs until the tissue surface is breached differs according to the tissue shape and stiffness. This topic has not been a focus of previous work. In the present paper, we propose a model with which to estimate the needle deflection that occurs until breaching of the tissue surface with consideration of the tissue shape and stiffness. This model comprises a cantilever beam supported by virtual springs that represent the interaction forces between the needle tip and tissue surface. The effects of different insertion angles and tissue stiffness on needle deflection are represented by changing the spring constants. The model was used in experiments involving four different insertion angles (0º, 15º, 30º, and 45º) and three different polyvinyl chloride (PVC) phantoms with different stiffness (100, 75, and 50%). We verified the proposed model with the 80% PVC phantom and showed a maximum error of 0.04 mm.

    KW - Deflection estimation

    KW - Mechanical-based model

    KW - Needle steering

    KW - Rayleigh–Ritz method

    UR - http://www.scopus.com/inward/record.url?scp=85052846162&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=85052846162&partnerID=8YFLogxK

    U2 - 10.1007/s40846-017-0359-5

    DO - 10.1007/s40846-017-0359-5

    M3 - Article

    AN - SCOPUS:85052846162

    VL - 38

    SP - 697

    EP - 706

    JO - Journal of Medical and Biological Engineering

    JF - Journal of Medical and Biological Engineering

    SN - 1609-0985

    IS - 5

    ER -