Optimal design of mri magnets with magnetic shielding

A. Ishiyama, M. Hondoh, N. Ishida, T. Onuki

Research output: Contribution to journalArticle

17 Citations (Scopus)

Abstract

We have been developing a novel optimal design method for a highly homogeneous superconducting coil system with magnetic shielding for Magnetic Resonance Imaging (MRI). The presented optimal design method is based on a combination of the hybrid finite - boundary element method for analysis of an axially symmetric nonlinear open boundary magnetic field problem and the mathematical programming method for solving the corresponding optimization problem. In this paper, the goal programming method and the nonlinear least squares method have been adopted. The relative merits of a combination of an active shielding and a room shielding have been investigated. The proposed design method can optimize both an MRI magnet and its magnetic shielding system simultaneously taking into account the maximum permissible intensity of stray field at a specified distance from the magnet center. Several design procedures and results of a I-tesla-magnet system are shown.

Original languageEnglish
Pages (from-to)1885-1888
Number of pages4
JournalIEEE Transactions on Magnetics
Volume25
Issue number2
DOIs
Publication statusPublished - 1989 Mar

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Electrical and Electronic Engineering

Fingerprint Dive into the research topics of 'Optimal design of mri magnets with magnetic shielding'. Together they form a unique fingerprint.

  • Cite this