Optimum angle of deflection electrodes of a cylindrical electrostatic analyzer

C. Oshima, R. Souda, M. Aono, Y. Ishizawa

    Research output: Contribution to journalArticle

    25 Citations (Scopus)

    Abstract

    The optimum deflection angle of the electrodes in a cylindrical electrostatic analyzer with real slits has been discussed on the basis of numerically calculated electron trajectories. The effect of fringing fields has been included exactly in this calculation. The optimum angle is 126.5°, being independent of the gap between the inner and outer electrodes. By this optimization, the first-order angle aberration can be eliminated completely.

    Original languageEnglish
    Pages (from-to)227-230
    Number of pages4
    JournalReview of Scientific Instruments
    Volume56
    Issue number2
    DOIs
    Publication statusPublished - 1985

    Fingerprint

    deflection
    Electrostatics
    analyzers
    electrostatics
    Electrodes
    electrodes
    Aberrations
    electron trajectories
    Trajectories
    slits
    Electrons
    aberration
    optimization

    ASJC Scopus subject areas

    • Instrumentation
    • Physics and Astronomy (miscellaneous)

    Cite this

    Optimum angle of deflection electrodes of a cylindrical electrostatic analyzer. / Oshima, C.; Souda, R.; Aono, M.; Ishizawa, Y.

    In: Review of Scientific Instruments, Vol. 56, No. 2, 1985, p. 227-230.

    Research output: Contribution to journalArticle

    Oshima, C. ; Souda, R. ; Aono, M. ; Ishizawa, Y. / Optimum angle of deflection electrodes of a cylindrical electrostatic analyzer. In: Review of Scientific Instruments. 1985 ; Vol. 56, No. 2. pp. 227-230.
    @article{2c34727776dd402d8de9451e68e99a7e,
    title = "Optimum angle of deflection electrodes of a cylindrical electrostatic analyzer",
    abstract = "The optimum deflection angle of the electrodes in a cylindrical electrostatic analyzer with real slits has been discussed on the basis of numerically calculated electron trajectories. The effect of fringing fields has been included exactly in this calculation. The optimum angle is 126.5°, being independent of the gap between the inner and outer electrodes. By this optimization, the first-order angle aberration can be eliminated completely.",
    author = "C. Oshima and R. Souda and M. Aono and Y. Ishizawa",
    year = "1985",
    doi = "10.1063/1.1138335",
    language = "English",
    volume = "56",
    pages = "227--230",
    journal = "Review of Scientific Instruments",
    issn = "0034-6748",
    publisher = "American Institute of Physics Publising LLC",
    number = "2",

    }

    TY - JOUR

    T1 - Optimum angle of deflection electrodes of a cylindrical electrostatic analyzer

    AU - Oshima, C.

    AU - Souda, R.

    AU - Aono, M.

    AU - Ishizawa, Y.

    PY - 1985

    Y1 - 1985

    N2 - The optimum deflection angle of the electrodes in a cylindrical electrostatic analyzer with real slits has been discussed on the basis of numerically calculated electron trajectories. The effect of fringing fields has been included exactly in this calculation. The optimum angle is 126.5°, being independent of the gap between the inner and outer electrodes. By this optimization, the first-order angle aberration can be eliminated completely.

    AB - The optimum deflection angle of the electrodes in a cylindrical electrostatic analyzer with real slits has been discussed on the basis of numerically calculated electron trajectories. The effect of fringing fields has been included exactly in this calculation. The optimum angle is 126.5°, being independent of the gap between the inner and outer electrodes. By this optimization, the first-order angle aberration can be eliminated completely.

    UR - http://www.scopus.com/inward/record.url?scp=0000461283&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=0000461283&partnerID=8YFLogxK

    U2 - 10.1063/1.1138335

    DO - 10.1063/1.1138335

    M3 - Article

    AN - SCOPUS:0000461283

    VL - 56

    SP - 227

    EP - 230

    JO - Review of Scientific Instruments

    JF - Review of Scientific Instruments

    SN - 0034-6748

    IS - 2

    ER -