Abstract
Based on the evaluation of oxygen transfer performance in terms of non-dimensional variables such as oxygen Sherwood number, fractional saturation and non-dimensional length, the conditions achieving the high performance of gas exchange is discussed. The non-dimensional length at the rated condition, in which the oxygen saturation at the outlet becomes 97%, was determined in the relation between the fractional saturation and non-dimensional length. Simultaneously the oxygen transfer performance expressed by oxygen Sherwood number has been evaluated in relation to the variation of non-dimensional length. Finally the design parameters such as internal diameter, tube length and residence time are determined based on the rated condition.
Original language | English |
---|---|
Pages (from-to) | 2383-2389 |
Number of pages | 7 |
Journal | Nippon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B |
Volume | 55 |
Issue number | 516 |
Publication status | Published - 1989 Aug |
ASJC Scopus subject areas
- Mechanical Engineering