Reliability of measurement of elastomer samples by soft tissue stiffness meter

Akira Takanashi, Hiroshi Karasuno, Munenori Katou, Ryo Konuma, Kotomi Shiota, Tadamitsu Matsuda, Shigeki Miyajima, Noboru Sakanoue, Masataka Hosoda, Junya Aizawa, Kouji Isosaki, Kouji Otsu

Research output: Contribution to journalArticle

Abstract

Purpose: We assessed the TK-03C soft tissue stiffness meter (STSM) (Tokushukeisoku Corporation, Japan) for inter/intra-rater reliability. This device measures load and displacement simultaneously. Subjects: Ten students from the Physical Therapy Department were the subjects. Methods: The 10 subjects measured the surface stiffness of three elastomer samples with known mechanical properties: a polyurethane test piece and two sorbos test pieces. They each took 7 consecutive readings using the STMS at the same site on each test piece. We calculated the displacement for a load of 10 N. Result: The overall rating reliability for all sites was high: site1, coefficient of variation (CV) (range)=5.3 (2.6-10.0); site2, CV=3.9 (2.3-6.1); site3, CV=5.0 (2.2-9.2). The intra-rater reliability of the 10 subjects was high showing an intra-class correlation coefficient (ICC) (1•1) ≥0.97. The inter-rater reliability was also high; ICC(2•1)=0.96. In addition, the hardness of the three test pieces were significantly different (p<0.01). Conclusion: It seems that this STSM is a reliable and useful device for objective quantitative evaluation of soft tissue stiffness.

Original languageEnglish
Pages (from-to)31-34
Number of pages4
JournalRigakuryoho Kagaku
Volume24
Issue number1
DOIs
Publication statusPublished - 2009
Externally publishedYes

Fingerprint

Elastomers
Hardness Tests
Equipment and Supplies
Polyurethanes
Reading
Japan
Students
Therapeutics

Keywords

  • Reliability
  • Soft tissue stiffness
  • Soft tissue stiffness meter

ASJC Scopus subject areas

  • Physical Therapy, Sports Therapy and Rehabilitation

Cite this

Takanashi, A., Karasuno, H., Katou, M., Konuma, R., Shiota, K., Matsuda, T., ... Otsu, K. (2009). Reliability of measurement of elastomer samples by soft tissue stiffness meter. Rigakuryoho Kagaku, 24(1), 31-34. https://doi.org/10.1589/rika.24.31

Reliability of measurement of elastomer samples by soft tissue stiffness meter. / Takanashi, Akira; Karasuno, Hiroshi; Katou, Munenori; Konuma, Ryo; Shiota, Kotomi; Matsuda, Tadamitsu; Miyajima, Shigeki; Sakanoue, Noboru; Hosoda, Masataka; Aizawa, Junya; Isosaki, Kouji; Otsu, Kouji.

In: Rigakuryoho Kagaku, Vol. 24, No. 1, 2009, p. 31-34.

Research output: Contribution to journalArticle

Takanashi, A, Karasuno, H, Katou, M, Konuma, R, Shiota, K, Matsuda, T, Miyajima, S, Sakanoue, N, Hosoda, M, Aizawa, J, Isosaki, K & Otsu, K 2009, 'Reliability of measurement of elastomer samples by soft tissue stiffness meter', Rigakuryoho Kagaku, vol. 24, no. 1, pp. 31-34. https://doi.org/10.1589/rika.24.31
Takanashi A, Karasuno H, Katou M, Konuma R, Shiota K, Matsuda T et al. Reliability of measurement of elastomer samples by soft tissue stiffness meter. Rigakuryoho Kagaku. 2009;24(1):31-34. https://doi.org/10.1589/rika.24.31
Takanashi, Akira ; Karasuno, Hiroshi ; Katou, Munenori ; Konuma, Ryo ; Shiota, Kotomi ; Matsuda, Tadamitsu ; Miyajima, Shigeki ; Sakanoue, Noboru ; Hosoda, Masataka ; Aizawa, Junya ; Isosaki, Kouji ; Otsu, Kouji. / Reliability of measurement of elastomer samples by soft tissue stiffness meter. In: Rigakuryoho Kagaku. 2009 ; Vol. 24, No. 1. pp. 31-34.
@article{d1a999f4ce2642cc9b9540456abe7e47,
title = "Reliability of measurement of elastomer samples by soft tissue stiffness meter",
abstract = "Purpose: We assessed the TK-03C soft tissue stiffness meter (STSM) (Tokushukeisoku Corporation, Japan) for inter/intra-rater reliability. This device measures load and displacement simultaneously. Subjects: Ten students from the Physical Therapy Department were the subjects. Methods: The 10 subjects measured the surface stiffness of three elastomer samples with known mechanical properties: a polyurethane test piece and two sorbos test pieces. They each took 7 consecutive readings using the STMS at the same site on each test piece. We calculated the displacement for a load of 10 N. Result: The overall rating reliability for all sites was high: site1, coefficient of variation (CV) (range)=5.3 (2.6-10.0); site2, CV=3.9 (2.3-6.1); site3, CV=5.0 (2.2-9.2). The intra-rater reliability of the 10 subjects was high showing an intra-class correlation coefficient (ICC) (1•1) ≥0.97. The inter-rater reliability was also high; ICC(2•1)=0.96. In addition, the hardness of the three test pieces were significantly different (p<0.01). Conclusion: It seems that this STSM is a reliable and useful device for objective quantitative evaluation of soft tissue stiffness.",
keywords = "Reliability, Soft tissue stiffness, Soft tissue stiffness meter",
author = "Akira Takanashi and Hiroshi Karasuno and Munenori Katou and Ryo Konuma and Kotomi Shiota and Tadamitsu Matsuda and Shigeki Miyajima and Noboru Sakanoue and Masataka Hosoda and Junya Aizawa and Kouji Isosaki and Kouji Otsu",
year = "2009",
doi = "10.1589/rika.24.31",
language = "English",
volume = "24",
pages = "31--34",
journal = "Rigakuryoho Kagaku",
issn = "1341-1667",
publisher = "Society of Physical Therapy Science (Rigaku Ryoho Kagakugakkai)",
number = "1",

}

TY - JOUR

T1 - Reliability of measurement of elastomer samples by soft tissue stiffness meter

AU - Takanashi, Akira

AU - Karasuno, Hiroshi

AU - Katou, Munenori

AU - Konuma, Ryo

AU - Shiota, Kotomi

AU - Matsuda, Tadamitsu

AU - Miyajima, Shigeki

AU - Sakanoue, Noboru

AU - Hosoda, Masataka

AU - Aizawa, Junya

AU - Isosaki, Kouji

AU - Otsu, Kouji

PY - 2009

Y1 - 2009

N2 - Purpose: We assessed the TK-03C soft tissue stiffness meter (STSM) (Tokushukeisoku Corporation, Japan) for inter/intra-rater reliability. This device measures load and displacement simultaneously. Subjects: Ten students from the Physical Therapy Department were the subjects. Methods: The 10 subjects measured the surface stiffness of three elastomer samples with known mechanical properties: a polyurethane test piece and two sorbos test pieces. They each took 7 consecutive readings using the STMS at the same site on each test piece. We calculated the displacement for a load of 10 N. Result: The overall rating reliability for all sites was high: site1, coefficient of variation (CV) (range)=5.3 (2.6-10.0); site2, CV=3.9 (2.3-6.1); site3, CV=5.0 (2.2-9.2). The intra-rater reliability of the 10 subjects was high showing an intra-class correlation coefficient (ICC) (1•1) ≥0.97. The inter-rater reliability was also high; ICC(2•1)=0.96. In addition, the hardness of the three test pieces were significantly different (p<0.01). Conclusion: It seems that this STSM is a reliable and useful device for objective quantitative evaluation of soft tissue stiffness.

AB - Purpose: We assessed the TK-03C soft tissue stiffness meter (STSM) (Tokushukeisoku Corporation, Japan) for inter/intra-rater reliability. This device measures load and displacement simultaneously. Subjects: Ten students from the Physical Therapy Department were the subjects. Methods: The 10 subjects measured the surface stiffness of three elastomer samples with known mechanical properties: a polyurethane test piece and two sorbos test pieces. They each took 7 consecutive readings using the STMS at the same site on each test piece. We calculated the displacement for a load of 10 N. Result: The overall rating reliability for all sites was high: site1, coefficient of variation (CV) (range)=5.3 (2.6-10.0); site2, CV=3.9 (2.3-6.1); site3, CV=5.0 (2.2-9.2). The intra-rater reliability of the 10 subjects was high showing an intra-class correlation coefficient (ICC) (1•1) ≥0.97. The inter-rater reliability was also high; ICC(2•1)=0.96. In addition, the hardness of the three test pieces were significantly different (p<0.01). Conclusion: It seems that this STSM is a reliable and useful device for objective quantitative evaluation of soft tissue stiffness.

KW - Reliability

KW - Soft tissue stiffness

KW - Soft tissue stiffness meter

UR - http://www.scopus.com/inward/record.url?scp=70049111591&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=70049111591&partnerID=8YFLogxK

U2 - 10.1589/rika.24.31

DO - 10.1589/rika.24.31

M3 - Article

AN - SCOPUS:70049111591

VL - 24

SP - 31

EP - 34

JO - Rigakuryoho Kagaku

JF - Rigakuryoho Kagaku

SN - 1341-1667

IS - 1

ER -