Side-scattered finger-photoplethysmography: Experimental investigations toward practical noninvasive measurement of blood glucose

Yasuhiro Yamakoshi, Kenta Matsumura, Takehiro Yamakoshi, Jihyoung Lee, Peter Rolfe, Yuji Kato, Koichi Shimizu, Ken Ichi Yamakoshi

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.

Original languageEnglish
Article number067001
JournalJournal of Biomedical Optics
Volume22
Issue number6
DOIs
Publication statusPublished - 2017 Jun 1
Externally publishedYes

Fingerprint

Photoplethysmography
glucose
blood
Glucose
Blood Glucose
Blood
Water absorption
Detectors
Wavelength
wavelengths
detectors
Radiation
water
circumferences
radiation detectors
Absorption spectra
Signal to noise ratio
radiation sources
Experiments
Tissue

Keywords

  • blood glucose absorption pulse glucometry
  • finger-photoplethysmography
  • laser diode
  • side-scattered radiation
  • water absorption

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Atomic and Molecular Physics, and Optics
  • Biomaterials
  • Biomedical Engineering

Cite this

Side-scattered finger-photoplethysmography : Experimental investigations toward practical noninvasive measurement of blood glucose. / Yamakoshi, Yasuhiro; Matsumura, Kenta; Yamakoshi, Takehiro; Lee, Jihyoung; Rolfe, Peter; Kato, Yuji; Shimizu, Koichi; Yamakoshi, Ken Ichi.

In: Journal of Biomedical Optics, Vol. 22, No. 6, 067001, 01.06.2017.

Research output: Contribution to journalArticle

Yamakoshi, Yasuhiro ; Matsumura, Kenta ; Yamakoshi, Takehiro ; Lee, Jihyoung ; Rolfe, Peter ; Kato, Yuji ; Shimizu, Koichi ; Yamakoshi, Ken Ichi. / Side-scattered finger-photoplethysmography : Experimental investigations toward practical noninvasive measurement of blood glucose. In: Journal of Biomedical Optics. 2017 ; Vol. 22, No. 6.
@article{c0383c7b6a5d4a66b82416199eaecb4b,
title = "Side-scattered finger-photoplethysmography: Experimental investigations toward practical noninvasive measurement of blood glucose",
abstract = "The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.",
keywords = "blood glucose absorption pulse glucometry, finger-photoplethysmography, laser diode, side-scattered radiation, water absorption",
author = "Yasuhiro Yamakoshi and Kenta Matsumura and Takehiro Yamakoshi and Jihyoung Lee and Peter Rolfe and Yuji Kato and Koichi Shimizu and Yamakoshi, {Ken Ichi}",
year = "2017",
month = "6",
day = "1",
doi = "10.1117/1.JBO.22.6.067001",
language = "English",
volume = "22",
journal = "Journal of Biomedical Optics",
issn = "1083-3668",
publisher = "SPIE",
number = "6",

}

TY - JOUR

T1 - Side-scattered finger-photoplethysmography

T2 - Experimental investigations toward practical noninvasive measurement of blood glucose

AU - Yamakoshi, Yasuhiro

AU - Matsumura, Kenta

AU - Yamakoshi, Takehiro

AU - Lee, Jihyoung

AU - Rolfe, Peter

AU - Kato, Yuji

AU - Shimizu, Koichi

AU - Yamakoshi, Ken Ichi

PY - 2017/6/1

Y1 - 2017/6/1

N2 - The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.

AB - The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.

KW - blood glucose absorption pulse glucometry

KW - finger-photoplethysmography

KW - laser diode

KW - side-scattered radiation

KW - water absorption

UR - http://www.scopus.com/inward/record.url?scp=85021643480&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85021643480&partnerID=8YFLogxK

U2 - 10.1117/1.JBO.22.6.067001

DO - 10.1117/1.JBO.22.6.067001

M3 - Article

C2 - 28636064

AN - SCOPUS:85021643480

VL - 22

JO - Journal of Biomedical Optics

JF - Journal of Biomedical Optics

SN - 1083-3668

IS - 6

M1 - 067001

ER -