T1 - Side-scattered finger-photoplethysmography
T2 - Experimental investigations toward practical noninvasive measurement of blood glucose
AU - Yamakoshi, Yasuhiro
AU - Matsumura, Kenta
AU - Yamakoshi, Takehiro
AU - Lee, Jihyoung
AU - Rolfe, Peter
AU - Kato, Yuji
AU - Shimizu, Koichi
AU - Yamakoshi, Ken Ichi
N1 - Publisher Copyright:
© The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Copyright:
Copyright 2017 Elsevier B.V., All rights reserved.
PY - 2017/6/1
Y1 - 2017/6/1
N2 - The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.
AB - The aim of this study was to discover a simple/convenient geometrical arrangement of radiation sources and detector to acquire finger-photoplethysmograms (PPGs) with wavelength regions of blood glucose (BGL) absorption, toward practical noninvasive BGL measurement. First, we compared PPGs with three wavelengths: 808 nm (without water absorption), 1160 nm (with weak water absorption), and 1600 nm (with nearly peak BGL absorption and strong water absorption), while the source-detector spacing was successively increased circumferentially around a fingertip. In 10 healthy subjects, we observed clear cardiac-related pulsatile components of PPG signals at 808 and 1160 nm in any incident positions with more than 15 dB of signal-to-noise ratio (S/N), but reliable PPG detections at 1600 nm with more than 10 dB of S/N was only possible when the source-detector distance was less than 3 mm around the fingertip circumference. Second, with this arrangement, an experiment was performed using six wavelengths to cover glucose absorption bands (from 1550 to 1749 nm), obtaining pulsatile PPG signals with more or less 15 dB of S/N. Through the present experiments, this orthogonal arrangement of the source and detector to detect forward- and side-scattered radiation through the tissue is appropriate for PPG measurements with wavelength regions where there is potential for BGL measurement.
KW - blood glucose absorption pulse glucometry
KW - finger-photoplethysmography
KW - laser diode
KW - side-scattered radiation
KW - water absorption
UR - http://www.scopus.com/inward/record.url?scp=85021643480&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85021643480&partnerID=8YFLogxK
U2 - 10.1117/1.JBO.22.6.067001
DO - 10.1117/1.JBO.22.6.067001
M3 - Article
C2 - 28636064
AN - SCOPUS:85021643480
SN - 1083-3668
VL - 22
JO - Journal of Biomedical Optics
JF - Journal of Biomedical Optics
IS - 6
M1 - 067001
ER -