抄録
This paper describes an optimal design of a highly homogeneous superconducting coil system for magnetic Resonance Imaging (MRI). The presented optimal design technique is based on a combination of a numerical Method for analysis of the magnetic field distribution and a mathematical programning method for solving the corresponding optimization problem. In this paper, the coil current density, the croos-sectional shape of coil, and the configuration of coil are optimizer so that a highly homogeneous magnetic field is realized.
本文言語 | English |
---|---|
ページ(範囲) | 922-925 |
ページ数 | 4 |
ジャーナル | IEEE Transactions on Magnetics |
巻 | 24 |
号 | 2 |
DOI | |
出版ステータス | Published - 1988 3 |
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Electrical and Electronic Engineering