Automatic stop of a microinjector distinctively in the cytosol or the vacuole of plant single-cells

Hideaki Matsuoka, Yohei Yamada, Shigetoshi Horikiri, Keisuke Hatta, Tomoaki Kubo, Yasuhiro Kushihashi, Mikako Saito, Yoshiyuki Miwa

    研究成果: Article査読

    抄録

    A 3-channel (3C) microinjector was prepared by pulling a glass capillary. One channel was used for a potential measuring electrode (MeaE) and the other two channels were used for electrophoretic introduction electrode (IntE) and its counter electrode (CE). The 3C microinjector was propelled by an oil pressure manipulator that was driven by a pulse motor, while the MeaE output was recorded continuously. When the 3C microinjector penetrated the cell membrane of a tobacco cultured cell, BY-2, the output potential changed sharply towards the negative direction. Then the microinjector progressed further into the cell. A marked change of the output potential occurred again, though it was towards the positive direction. These changes were speculated as the passages of cell membrane and vacuolar membrane, respectively. Using the differential of these changes as the stop signal, the microinjector could rightly be stopped either in the cytosol or the vacuole, which was confirmed by the dye diffusion pattern.

    本文言語English
    ページ(範囲)513-517
    ページ数5
    ジャーナルElectrochemistry
    75
    7
    出版ステータスPublished - 2007 7

    ASJC Scopus subject areas

    • 電気化学

    フィンガープリント

    「Automatic stop of a microinjector distinctively in the cytosol or the vacuole of plant single-cells」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

    引用スタイル