抄録
A novel digital PET scanner based on Time over Threshold method is developed. The positron emission tomography (PET) is composed of 144channel Ce:Gd3Al2Ga3O12 (GAGG)-Avaranche photodiode (APD) detector arrays individually coupled with custom designed Time over Threshold (ToT) application-specific integrated circuit (ASIC) to realize the high count rate and good spatial resolution. Such an imaging system provides a simple front-end circuit and flexible digital signal processing like multiplexing such as a pulse train method. The measured energy resolution of the detector system was 6.7% for the 511 keV peak, and 4.25 ns time resolution was measured with a single detector module. The measured spatial resolution for a point source was 1.37 mm FWHM for our initial data with a columnar 22Na source.
本文言語 | English |
---|---|
論文番号 | 6722988 |
ページ(範囲) | 348-352 |
ページ数 | 5 |
ジャーナル | IEEE Transactions on Nuclear Science |
巻 | 61 |
号 | 1 |
DOI | |
出版ステータス | Published - 2014 2月 |
ASJC Scopus subject areas
- 核物理学および高エネルギー物理学
- 原子力エネルギーおよび原子力工学
- 電子工学および電気工学