抄録
Hydrogen absorption of biomedical titanium and Ni-Ti alloys in a neutral fluoride (2.0% NaF) solution for up to 10,000 h at 37°C has been evaluated by means of hydrogen thermal desorption analysis. For α titanium (commercial pure titanium), the amount of absorbed hydrogen was, at most, 10-30 mass ppm, and the corrosion product and hydride formation were revealed on the surface of the specimen by X-ray diffraction analysis. Ni-Ti superelastic alloy absorbed ∼150 mass ppm of hydrogen, which was probably sufficient to result in the pronounced degradation of the mechanical properties, although corrosion was hardly observed. In contrast, hydrogen absorption of α-β titanium (Ti-6Al-4V) and β titanium (Ti-11.3Mo-6.6Zr-4.3Sn) alloys was negligible, although general corrosion was observed. The results of the present study indicate that the susceptibility of titanium and Ni-Ti alloys to hydrogen absorption in the neutral fluoride solution is different from that in the acidic fluoride solution reported previously.
本文言語 | English |
---|---|
ページ(範囲) | 204-210 |
ページ数 | 7 |
ジャーナル | Journal of Biomedical Materials Research - Part B Applied Biomaterials |
巻 | 78 |
号 | 1 |
DOI | |
出版ステータス | Published - 2006 7月 |
ASJC Scopus subject areas
- 生体医工学
- 生体材料