Improvement of particle alignment control and precise image acquisition for on-chip high-speed imaging cell sorter

Akihiro Hattori, Tomoyuki Kaneko, Kenji Yasuda

研究成果: Article

8 被引用数 (Scopus)

抄録

We have developed a real-time imaging cell sorting system composed of a micrometer-sized gel-electrode-embedded microfluidic sorting chip and a real-time image analysis/recognition unit equipped with a high-speed camera and image processing circuits. For the microfluidic continuous cell sorting, we have examined the precise position and velocity control of flowing particles and the precise acquisition of microscopic images of flowing particles. The results showed that (1) hydrodynamic focusing can line up particles precisely within a range of 5 μm particle size distribution, (2) active air pressure-driven flow velocity control can create the flow in the microfluidic pathways up to 160mm/s with 0.15MPa air pressure maintaining linear correlation between air pressure and flow velocity, and (3) 1μs flash illumination can prevent the blur even under 200mm/s flow. Applying the above elements into the system, the recognition error of target particles was within 5% for 2 μm particles with 2.5mm/s flow. The experimental results demonstrate the potential of the image index-based on-chip cell sorter for practical application.

本文言語English
論文番号06GL06
ジャーナルJapanese journal of applied physics
50
6 PART 2
DOI
出版ステータスPublished - 2011 6 1
外部発表はい

ASJC Scopus subject areas

  • Engineering(all)
  • Physics and Astronomy(all)

フィンガープリント 「Improvement of particle alignment control and precise image acquisition for on-chip high-speed imaging cell sorter」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。