Improvement of the CdTe diode detectors using a guard-ring electrode

Kazuhiro Nakazawa*, Kousuke Oonuki, Takaaki Tanaka, Yoshihito Kobayashi, Ken'ichi Tamura, Takefumi Mitani, Goro Sato, Shin Watanabe, Tadayuki Takahashi, Ryoichi Ohno, Akihito Kitajima, Yoshikatsu Kuroda, Mitsunobu Onishi

*この研究の対応する著者

研究成果: Article査読

抄録

Recent results from the CdTe Schottky diode detectors employing a guard-ring electrode are reported. Cathode electrode, made of Pt, was separated into an active electrode(s) and a surrounding guard-ring. Typical leakage current of a device with an active area of 2 × 2 mm2 and 0.5 mm thickness surrounded by a guard-ring, is 7 pA and 20 pA at a bias of 100 V and 500 V, respectively, operated at 20°C. Spectral resolution of this device is 0.93 keV and 1.2 keV (FWHM) at 59.5 keV and 122 keV, respectively, operated at 20°C with a bias of 800 V. Detailed study of the characteristics of these devices working as a gamma-ray detector is presented.

本文言語English
論文番号R16-4
ページ(範囲)3583-3587
ページ数5
ジャーナルUnknown Journal
5
出版ステータスPublished - 2003