Improvements in time resolution and signal-to-noise ratio in a compact pico-second pulse radiolysis system

Hiroyuki Nagai, Masaaki Kawaguchi, Kazuyuki Sakaue, Keita Komiya, Tomoaki Nomoto, Yoshio Kamiya, Yoshimasa Hama, Masakazu Washio, Kiminori Ushida, Shigeru Kashiwagi, Ryunosuke Kuroda

研究成果: Article査読

6 被引用数 (Scopus)

抄録

A compact pico-second pulse radiolysis system has been developing at Waseda University for studying primary processes in radiation chemistry. The system is composed of a photo-injector system and a pico-second all-solid-state laser system. An infrared (IR) and an ultraviolet (UV) laser pulses are obtained from mode-locked Nd:YLF laser system and used for generation of the white light continuum as a probe light and the irradiation to the Cu cathode of a photo-cathode RF-gun, respectively. To improve signal-to-noise (S/N) ratio and time resolution of this pulse radiolysis system, we optimized both probe light and pump electron beam. As a result, our pico-second pulse radiolysis system has been enough to study the primary processes of radiation chemistry. The experimental results and the improvements of our system are described in this paper.

本文言語English
ページ(範囲)82-86
ページ数5
ジャーナルNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
265
1
DOI
出版ステータスPublished - 2007 12