Long-Term Durability Test for the Left Ventricular Assist System EVAHEART under the Physiologic Pulsatile Load

Tomoya Kitano*, Kiyotaka Iwasaki

*この研究の対応する著者

    研究成果: Article査読

    3 被引用数 (Scopus)

    抄録

    The EVAHEART Left Ventricular Assist System (LVAS) was designed for the long-term support of a patient with severe heart failure. It has an original water lubrication system for seal and bearing and wear on these parts was considered one of its critical failure modes. A durability test focusing on wear was designed herein. We developed a mock loop, which generates a physiologic pulsatile flow and is sufficiently durable for a long-term test. The pulsatile load and the low fluid viscosity enable the creation of a severe condition for the mechanical seal. A total of 18 EVAHEART blood pumps completed 2 years of operation under the pulsatile condition without any failure. It indicated the EVAHEART blood pump had a greater than 90% reliability with a 88% confidence level. The test was continued with six blood pumps and achieved an average of 8.6 years, which was longer than the longest clinical use in Japan. The test result showed that no catastrophic, critical, marginal, or minor failures of the blood pump or their symptoms were observed. The seal performance was maintained after the test. Moreover, the surface roughness did not change, which showed any burn or abnormal wear occurred. The original water lubrication system equipped in EVAHEART LVAS prevent severe wear on the seal and the bearing, and it can be used in the bridge to transplant and destination therapy.

    本文言語English
    ページ(範囲)168-174
    ページ数7
    ジャーナルASAIO Journal
    DOI
    出版ステータスAccepted/In press - 2018 1 1

    ASJC Scopus subject areas

    • 生物理学
    • バイオエンジニアリング
    • 生体材料
    • 生体医工学

    フィンガープリント

    「Long-Term Durability Test for the Left Ventricular Assist System EVAHEART under the Physiologic Pulsatile Load」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

    引用スタイル