Multiple purpose simulator using a natural porcine mitral valve

Makoto Arita, Sumihiro Tono, Hitoshi Kasegawa, Mitsuo Umezu

研究成果: Article査読

8 被引用数 (Scopus)

抄録

An in vitro pulsatile simulator with a porcine mitral valve was developed in order to simulate physiologic and diseased mitral valve conditions. Evaluation of these conditions was conducted from a hydrodynamic and annulus behavior point of view. We found it possible to simulate mild "mitral valve prolapse" and to obtain quantitative data related to the condition. The diseased condition produced a 40% greater regurgitant volume than that observed under the normal condition (p < 0.0001). Regarding the leakage volume, the diseased condition exhibited about 2.6 times more leakage than the normal condition. The mitral valve simulator proposed in this study is considered fairly stable with respect to both hemodynamics and the behavior of the annulus, and it is an adequate simulator for modeling various types of normal and diseased mitral valve conditions.

本文言語English
ページ(範囲)350-356
ページ数7
ジャーナルAsian Cardiovascular and Thoracic Annals
12
4
DOI
出版ステータスPublished - 2004 12
外部発表はい

ASJC Scopus subject areas

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

フィンガープリント 「Multiple purpose simulator using a natural porcine mitral valve」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル