抄録
In a positron-sensitive probe composed of a plastic scintillator and a bismuth germanate (BGO), scattered annihilation photons in the plastic scintillator become background counts. Although these scattered annihilation photons can be rejected by higher threshold level settings for the scintillation pulse of the plastic scintillator and for that of the BGO, the system sensitivity is reduced. We have theoretically and experimentally optimized the threshold levels for both the plastic scintillator and the BGO. After calculating the energy loss in the plastic scintillator and the BGO for the scattered annihilation photons, we measured the background counts of a positron-sensitive probe by changing these threshold levels. Results revealed that one optimum threshold setting of the positron-sensitive probe was 0.3 of the peak level of the pulse for the plastic scintillator and 0.7 of that for the BGO. With these threshold levels, the background counts could be decreased to less than 0.2% of the true positron counts.
本文言語 | English |
---|---|
ページ(範囲) | 251-256 |
ページ数 | 6 |
ジャーナル | Annals of Nuclear Medicine |
巻 | 18 |
号 | 3 |
DOI | |
出版ステータス | Published - 2004 5月 |
外部発表 | はい |
ASJC Scopus subject areas
- 放射線学、核医学およびイメージング