Takehiko Fujiwara*, Shoichi Watanuki, Seiichi Yamamoto, Masayasu Miyake, Shinya Seo, Masatoshi Itoh, Keizou Ishii, Hikonojyou Orihara, Hiroshi Fukuda, Tomohiko Satoh, Keishi Kitamura, Kazumi Tanaka, Shigekazu Takahashi

*この研究の対応する著者

研究成果: Article査読

108 被引用数 (Scopus)

抄録

The SET-2400W is a newly designed whole-body PET scanner with a large axial field of view (20 cm). Its physical performance was investigated and evaluated. The scanner consists of four rings of 112 BGO detector units (22.8 mm in-plane x 50 mm axial x 30 mm depth). Each detector unit has a 6 (in- plane) x 8 (axial) matrix of BGO crystals coupled to two dual photomultiplier tubes. They are arranged in 32 rings giving 63 two-dimensional image planes. Sensitivity for a 20-cm cylindrical phantom was 6.1 kcps/kBq/ml (224 kcps/μCi/ml) in the 2D clinical mode, and to 48.6 kcps/kBq/ ml (1.8 Mcps/μCi/ml) in the 3D mode after scatter correction. In-plane spatial resolution was 3.9 mm FWHM at the center of the field-of-view, and 4.4 mm FWHM tangentially, and 5.4 mm FWHM radially at 100 mm from the center. Average axial resolution was 4.5 mm FWHM at the center and 5.8 mm FWHM at a radial position 100 mm from the center. Average scatter fraction was 8% for the 2D mode and 40% for the 3D mode. The maximum count rate was 230 kcps in the 2D mode and 350 kcps in the 3D mode. Clinical images demonstrate the utility of an enlarged axial field-of-view scanner in brain study and whole- body PET imaging.

本文言語English
ページ(範囲)307-313
ページ数7
ジャーナルAnnals of Nuclear Medicine
11
4
DOI
出版ステータスPublished - 1997
外部発表はい

ASJC Scopus subject areas

  • 放射線学、核医学およびイメージング

フィンガープリント

「Performance evaluation of a large axial field-of-view PET scanner: SET- 2400W」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル