抄録
High performance multi reagent injection system that enables selection and injection four kinds of reagents to arbitrary arrayed 16 microwells is presented. Integrated three microvalves and a reagent purge line to each microwell realizes precise control of the injected reagent volume in 1 nl order and negligible residual volume less than 0.05 nl. 60 independent actuated microvalves for controlling the reagent selector and the microwells are realized by only 16 outer pneumatic control lines applying addressable valve control system.
元の言語 | English |
---|---|
ホスト出版物のタイトル | Proceedings of Conference, MicroTAS 2009 - The 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences |
出版者 | Chemical and Biological Microsystems Society |
ページ | 1488-1490 |
ページ数 | 3 |
ISBN(印刷物) | 9780979806421 |
出版物ステータス | Published - 2009 |
イベント | 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2009 - Jeju 継続期間: 2009 11 1 → 2009 11 5 |
Other
Other | 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2009 |
---|---|
市 | Jeju |
期間 | 09/11/1 → 09/11/5 |
Fingerprint
ASJC Scopus subject areas
- Chemical Engineering (miscellaneous)
- Bioengineering
これを引用
Precise volume controlled multireagents injective microwell array for efficient cell function analysis. / Shibata, Yoko; Kawai, Kentaro; Kanai, Masaki; Shoji, Shuichi.
Proceedings of Conference, MicroTAS 2009 - The 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences. Chemical and Biological Microsystems Society, 2009. p. 1488-1490.研究成果: Conference contribution
}
TY - GEN
T1 - Precise volume controlled multireagents injective microwell array for efficient cell function analysis
AU - Shibata, Yoko
AU - Kawai, Kentaro
AU - Kanai, Masaki
AU - Shoji, Shuichi
PY - 2009
Y1 - 2009
N2 - High performance multi reagent injection system that enables selection and injection four kinds of reagents to arbitrary arrayed 16 microwells is presented. Integrated three microvalves and a reagent purge line to each microwell realizes precise control of the injected reagent volume in 1 nl order and negligible residual volume less than 0.05 nl. 60 independent actuated microvalves for controlling the reagent selector and the microwells are realized by only 16 outer pneumatic control lines applying addressable valve control system.
AB - High performance multi reagent injection system that enables selection and injection four kinds of reagents to arbitrary arrayed 16 microwells is presented. Integrated three microvalves and a reagent purge line to each microwell realizes precise control of the injected reagent volume in 1 nl order and negligible residual volume less than 0.05 nl. 60 independent actuated microvalves for controlling the reagent selector and the microwells are realized by only 16 outer pneumatic control lines applying addressable valve control system.
KW - Microwell array
KW - Precise volume injection
KW - Valve control system
UR - http://www.scopus.com/inward/record.url?scp=80052106897&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=80052106897&partnerID=8YFLogxK
M3 - Conference contribution
AN - SCOPUS:80052106897
SN - 9780979806421
SP - 1488
EP - 1490
BT - Proceedings of Conference, MicroTAS 2009 - The 13th International Conference on Miniaturized Systems for Chemistry and Life Sciences
PB - Chemical and Biological Microsystems Society
ER -