抄録
The laser-induced breakdown spectroscopy (LIBS) using micro-droplet sodium chloride solution is presented. Micronizing the sample made possible that whole volume of a sample was confined in the laser beam spot area and was separated from its surrounding condition. If the sample's physical state was liquid, the density of solution could be controlled as needed. Originally designed ink-jet system for sampling procedure was presented. According to the new method, improved calibration curves for the LIBS quantitative measurement were obtained.
本文言語 | English |
---|---|
ホスト出版物のタイトル | Proceedings of IEEE Sensors |
ページ | 291-294 |
ページ数 | 4 |
DOI | |
出版ステータス | Published - 2006 |
イベント | 2006 5th IEEE Conference on Sensors - Daegu 継続期間: 2006 10月 22 → 2006 10月 25 |
Other
Other | 2006 5th IEEE Conference on Sensors |
---|---|
City | Daegu |
Period | 06/10/22 → 06/10/25 |
ASJC Scopus subject areas
- 工学(その他)
- 電子工学および電気工学