Surface hydrophilic treatment of polyurea film realized by vacuum ultraviolet light irradiation and its application for poly(methylmethacrylate) blood analysis chip

H. Shinohara*, Y. Takahashi, J. Mizuno, S. Shoji

*この研究の対応する著者

研究成果: Article査読

15 被引用数 (Scopus)

抄録

A room temperature hydrophilic method of the polyurea film surface was developed using vacuum ultraviolet (VUV) light at 172 nm in wavelength. The treated film has highly hydrophilic surface for long time (∼2 months). The stable hydrophilic surface, water contact angle of smaller than 20°, was realized. It is kept even after very hard condition of ultrasonic cleaning in de-ionized water. The polyurea film was applied for poly(methylmethacrylate) (PMMA) blood analysis chip. The performance of the surface treatment was evaluated by actual human whole blood flow. The new chip can reduce the adhesion of platelets and white blood cells, even after ultrasonic cleaning in surfactant-induced water.

本文言語English
ページ(範囲)374-379
ページ数6
ジャーナルSensors and Actuators, B: Chemical
132
2
DOI
出版ステータスPublished - 2008 6月 16

ASJC Scopus subject areas

  • 電子材料、光学材料、および磁性材料
  • 器械工学
  • 凝縮系物理学
  • 表面、皮膜および薄膜
  • 金属および合金
  • 電子工学および電気工学
  • 材料化学