ULTRASONIC BEAMWIDTH EFFECT ON EVALUATION OF ATTENUATION BY SAMPLING MEDICAL B-MODE IMAGING DATA.

Kawan Soetanto*, Shigeo Ohtsuki, Motoyoshi Okujima, Motonao Tanaka

*この研究の対応する著者

研究成果: Conference article査読

抄録

A technique has been previously proposed for evaluating the ultrasonic attenuation constant of scattering media such as human liver. The technique is based on processing the data sampled from the ultrasonic B-mode images. The relation between the lateral sampling interval and the ultrasonic beamwidth in a given distance is investigated. Experiments with a tissue phantom compounded of fine glass balls and agar indicate that the appropriate sampling interval is half of the minus 3-dB beamwidth.

本文言語English
ページ(範囲)1773-1776
ページ数4
ジャーナルICASSP, IEEE International Conference on Acoustics, Speech and Signal Processing - Proceedings
出版ステータスPublished - 1986 12 1
外部発表はい